Development and Evaluation of an Air-Curtain Fume Cabinet with Considerations of its Aerodynamics
1 Department of Mechanical Engineering, National Taiwan University of Science and Technology 43 Keelung Road, Section 4, Taipei, Taiwan, Republic of China
2 Institute of Occupational Medicine and Industrial Hygiene, National Taiwan University 1 Jen-Ai Road, Section 1, Taipei, Taiwan, Republic of China
3 Institute of Occupational Safety and Health, Council of Labor Affairs, Executive Yuan No. 99, Lane 407, Hengke Road, Shijr City, Taipei County, Taiwan, Republic of China
4 Graduate Institute of Environmental Health, College of Public Health China Medical University, 91 Hsueh-Shih Road, Taichung, Taiwan, Republic of China
*Author to whom correspondence should be addressed. Tel: +886 2 2737 6488; fax: +886 2 2737 6460; e-mail: rfhuang{at}mail.ntust.edu.tw
In order to avoid the inherent aerodynamic difficulties of the conventional fume hood, an innovative designthe air curtain-isolated fume hood is developed. The new hood applies a specially designed air curtain (which is generated by a narrow planar jet and a suction slot flow at low velocities) across the sash plane. The hood constructed for the study is full size and transparent for flow visualization. The aerodynamic characteristics are diagnosed by using the laser-light-sheet-assisted smoke flow visualization method. Four characteristic air-curtain flow modes are identified in the domain of jet and suction velocities when the sash remains static. Some of these characteristic flow modes have much improved flow patterns when compared with those of the conventional fume hoods. From the viewpoint of the aerodynamics and mass transport, the results indicate that the air curtain properly setup across the sash opening allows almost no sensible exchange of momentum and mass between the flowfields of the cabinet and the outside environment. Two standard sulfur hexafluoride (SF6) tracer gas concentration measurement methods following the ANSI/ASHRAE 110-1995 standard and the prEN14175 protocol for static test are employed to examine the contaminant leakage levels. Results of the rigorous examinations of leakage show unusually satisfactory hood performance. The leakage of the tracer gas can approach almost null (<0.001 p.p.m.) if the jet and suction velocities are properly adjusted.
Keywords: air-curtain fume hood aerodynamic characteristics fume cupboard fume hood
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
S.-J. Tsai, R. F. Huang, and M. J. Ellenbecker Airborne Nanoparticle Exposures while Using Constant-Flow, Constant-Velocity, and Air-Curtain-Isolated Fume Hoods Ann. Hyg., November 20, 2009; (2009) mep074v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. F. Huang and C. I. Chou Flow and Performance of an Air-curtain Biological Safety Cabinet Ann. Hyg., June 1, 2009; 53(4): 425 - 440. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.-J. CHERN and W.-Y. CHENG Numerical Investigation of Turbulent Diffusion in Push Pull and Exhaust Fume Cupboards Ann. Hyg., August 1, 2007; 51(6): 517 - 531. [Abstract] [Full Text] [PDF] |
||||