Skip Navigation

This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by POIROT, P.
Right arrow Articles by HÉRY, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by POIROT, P.
Right arrow Articles by HÉRY, M.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Ann. occup. Hyg., Vol. 48, No. 1, pp. 75-84, 2004
© 2004 British Occupational Hygiene Society
Published by Oxford University Press

Determination of Short-term Exposure with a Direct Reading Photoionization Detector

P. POIROT1,*, I. SUBRA1, F. GÉRARDIN1, V. BAUDIN2, S. GROSSMANN3 and M. HÉRY4

1 Département Ingénierie des Procédés; 2 Service Epidémiologie en Entreprise; 3 Département Polluant et Santé; 4 Direction Scientifique, INRS, BP 27, F-54501 Vandoeuvre Cedex, France

Received 21 October 2002; in final form 12 June 2003

The use of a direct reading photoionization detector (PID) to determine short-term solvent exposures is described in the present paper. To assess the relevance of such a total exposure evaluation it was necessary to compare it with the real concentration of pollutants. This comparison was made by measuring in parallel with the PID determination the concentration of each pollutant using a standard technique, i.e. sampling on charcoal tubes and subsequent analysis by gas chromatography. Laboratory tests showed that the linearity of the answer of the PID is good for many compounds and for a mixture of these compounds. Similar tests were carried out for painters in workplaces with the same good correlations (determination coefficient r2 close to 1) between the PID response and the real concentration of the pollutants measured on the sampling tubes. The use of PID also allowed determination of the exposure profile of the workers and comparison of the short-term exposure to the corresponding limit values. Many cases of the short-term limit values being exceeded were revealed by use of the PID, although very few cases of the long-term limit values have been found by the usual sampling (charcoal tube) and analytical (gas chromatography) methods.

Keywords: photoionization detector; short-term exposure; solvent; limit values


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
ANN OCCUP HYGHome page
E. Langlois
GABIE and Perkin Elmer Passive Sampler Performance under Fluctuating Concentration Conditions
Ann. Hyg., June 1, 2008; 52(4): 239 - 247.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.