Ann. occup. Hyg., Vol. 47, No. 4, pp. 305-312, 2003
© 2003 British Occupational Hygiene Society
Published by Oxford University Press
Permeation of 70% Isopropyl Alcohol Through Surgical Gloves: Comparison of the Standard Methods ASTM F739 and EN 374

Department of Industrial Hygiene and Toxicology, Finnish Institute of Occupational Health, Topeliuksenkatu 41 a A, FIN-00250 Helsinki, Finland
Received 1 July 2002; in final form 17 December 2002
Standard test methods ASTM F739 and EN 374 were compared by assessing the permeation of 70% isopropyl alcohol (2-propanol) through seven brands of surgical gloves. The two standards differ in the flow rates of the collection medium and in the chemical permeation rate at which the breakthrough time (BTT) is detected, the EN detection level being 10 times higher than the permeation rate used by ASTM. In a departure from the EN standard method, a 4 h testing time was used instead of 8 h. All of the tested gloves were from the same manufacturer and were made from either natural rubber (NR) (six brands) or chloroprene rubber (CR) (one brand). Two of the NR glove brands were double layered. For the thin NR gloves (0.22, 0.28 and 0.27 mm) the permeation rates were higher throughout the tests with a flow rate of 474 ml/min (EN) of the collection medium (nitrogen) compared with the permeation rates obtained with a flow rate of 52 ml/min (ASTM). These resulted in BTTs of 4.6, 6.5 and 7.6 min (EN) and 4.8, 6.5 and 9.1 min (ASTM), respectively. No statistical difference could be observed between the BTT values obtained with the two standard methods for any of the thin gloves. Thus, although the ASTM standard has a lower criterion for the detection of permeation, it does not necessarily produce shorter BTTs. For the better barriers the methods yielded more equivalent permeation rate curves and thus the EN BTTs were longer than the ASTM BTTs: the EN results were 21, 80, 122 and >240 min compared with the ASTM results of 12, 32, 38 and 103 min for glove thicknesses of 0.37 (NR), 0.22 + 0.22 (double layered NR), 0.31 + 0.29 (double layered NR) and 0.19 mm (CR), respectively.
Keywords: gloves; isopropyl alcohol; permeation; 2-propanol; standards; surgical
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
G. Korinth, K. Schmid, O. Midasch, M. I. Boettcher, J. Angerer, and H. Drexler Investigations on Permeation of Mitomycin C Through Double Layers of Natural Rubber Gloves Ann. Hyg., October 5, 2007; (2007) mem039v1. [Abstract] [Full Text] [PDF] |
||||